Clinical Aspects
Bhavani Shankar Kodali MD
Analysis of CO2 waveforms:
Elevated inspiratory base line:
Normal | Rebreathing -Exhausted soda line – Elevation of the base line |
Trend | Waveform |
Contamination of CO2 monitor (sudden elevation of base line and top line) | Contamination of CO2 monitor – trend |
Inspiratory valve malfunction – Elevation of the base line, prolongation of down stroke, prolongation of phase III |
Capnogram during the use of Bain anesthetic circuit. Inspiratory base line and phase I are elevated above the zero due to rebreathing. Note the rebreathing wave during inspiration. Inspiration = |
References:
2. Berman LS, Pyles ST. Capnographic detection of anaesthesia circle valve malfunctions. Can J Anaesh 1988;35:473-5.
3. Van Ganderingen HR, Gravenstein N, Van der Aa JJ, Gravenstein JS. Computer-assisted capnogram analysis. J Clin Monit 1987;3:194-200.
4. Weingarten M. Respiratory monitoring of carbon dioxide and oxygen: a ten-year perspective. J Clin Monit 1990;6:217-25.
5. Kumar AY, Bhavani-Shankar K, Moseley HSL, Delph Y. Inspiratory valve malfunction in a circle system: Pitfalls in capnography. Can J Anaesth 1992;39:997-9.
6. Bhavani-Shankar K, Moseley H, Kumar AY, Delph Y. Anaesthesia and capnometry. (Review article). Can J Anaesth 1992;39:617-32.
7. Adams AP. Capnography and pulse oximetery. In: Atkins RS, Adams AP (Eds.). Recent Advances in Anaesthesia and Analgesia. London: Churchill Livingston, 1989;155-75.
8. Sweadlow DB, Irving SM. Monitoring and patient safety. In: Blitt CD (Ed.). Monitoring in Anesthesia and Critical Care Medicine. 2nd ed., New York: Churchill Livingstone, 1990;33:64.