Capnography in Emergency Tracheal Intubations

 

Capnography in Emergency Tracheal Intubations

Bhavani Shankar Kodali MD

 

Majority of emergency intubations are performed in surroundings that are less than ideal circumstances as compared to those in operating rooms. Therefore all tracheal intubations should be confirmed for endotracheal placement. Esophageal intubation or misplaced endotracheal tube can result in unsuccessful resuscitation, or can result in worsening of cardio respiratory status.

Until recently, calorimetric devices are the only portable devices suited for this purpose. Calorimetric devices have inherent false positives and false negatives as described below. There are some battery rechargeable portable capnographs for this purpose. A major disadvantage of these modules have been warming up time required before being ready for use. However, recently a small portable quantitative capnometer, developed and manufactured in Sweden, with 0.5 second response time has been introduced. This capnometer is marketed in the US as the Capnocheck® and as the EMMATM worldwide. It is described below. These units are new addition to capnography armamentarium that will encourage emergency care providers to confirm endotracheal intubation reliably. Possible applications are during field intubations, emergency room intubation and respiratory monitoring, hospital floor codes, hospital floor intubations, and intubations and respiratory monitoring in post-anesthesia care units. Intensive care units that have not realized full benefits of capnography modules can use these portable units for confirming endotracheal intubations.

 

Chemical method of CO2 measurement

chemical.gif

A pH-sensitive chemical indicator is enclosed in a plastic housing and is connected to the gas stream between the endotracheal tube and the anesthesia circuit. The pH sensitive indicator changes color when exposed to CO2.1 The color varies between expiration and inspiration, as inspiration, as level increases or decreases. The color changes from purple (when exposed to room air or oxygen) to yellow (when exposed to 4% CO2). The response time of the device is sufficiently fast to detect changes of CO2 breath-by breath.1 However, this device is not very sensitive when CO2 output is low as is during CPR. Easy cap II is an example of such pH sensitive indicator devices.

False negative results may occur even with correct endotracheal tube placement in patients in cardiac arrest, in whom sufficient CO2 may not be present in the lungs. It is also more relevant to point out the possibility of color change in the device due to agents other than exhaled carbon dioxide (false positive results). Gastric contents, mucus, and drugs such as epinephrine can cause false positive results. It is imperative that clinicians using these devices be aware of this limitation. One way to avoid this pitfall is to observe the change in color in the device with each breath. A false positive result causes a permanent color change in the device; hence, the color does not vary with ventilation.2,3

Reference:

1.O'Flaherty D, Adams AP. The end-tidal carbon dioxide detector. Assessment of new method to distinguish oesophageal from tracheal intubation. Anaesthesia 1990;45:653-5.

2.Srinivasa V, Kodali BS. Video on orotrachal intubation. NEJM 2007;357:6:620

3.Srinivasa V, Kodali BS. Caution when using colorimetry to confirm endotracheal intubation. Anesth Analg 2007;104:738

 

EMMATM Digital Capnometer

capnocheck.jpg

This is a small portable pocket size device that quantitatively measures CO2 concentrations and respiratory rates, when interposed across the airway. This device uses a miniaturized mainstream infra red ETCO2 technology providing CO2 measurements. Unlike other capnometers, the EMMATM / Capnocheck® Capnometer does not need to be calibrated. The unit can be set to alarm when the ETCO2 is out of set range. The measuring range is 0-99 mmHg with an accuracy of 0-40 mmHg ± 2 mmHg, 41-99 mmHg 6%. System response time is less than 0.5 seconds. The warm up time is less than 5 seconds. The capnometer is so light weight that it can be hung around the neck and carried all the time by emergency personnel. The device operates on two AAA batteries for eight hours. Although there are no formal publications using this device for emergency use, given the well authenticated application of capnography in emergent intubations, the device should be very useful.

For details of operation, click here

For further details, click here. Close the window to return back to this site again

Disclaimers Legal Notice


Disclaimers

Copyright Protection

Capnography.com. All rights reserved. Reproduction in whole or in part in any form or medium without express written permission of Bhavani Shankar Kodali is prohibited.

Disclaimer for Commercial Supporters

Capnography.com does not have any commercial supporters, either financial, or technical, for producing and maintaining this site. Therefore the commercial supporters have no input as to material included or excluded in the content and syllabus used in the educational website.

Errors and Omissions

THIS SITE IS PROVIDED BY Bhavani Shankar Kodali (Capnography.com). ON AN "AS IS" AND "AS AVAILABLE" BASIS. Capnography.com MAKES NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO THE OPERATION OF THE SITE, THE INFORMATION, CONTENT OR MATERIALS INCLUDED ON THIS SITE. TO THE FULL EXTENT PERMISSIBLE BY APPLICABLE LAW, Bhavani Shankar Kodali of Capnography.com DISCLAIMS ALL WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Bhavani Shankar Kodali of capnography.com WILL NOT BE LIABLE FOR ANY DAMAGES OF ANY KIND ARISING FROM THE USE OF THIS SITE, INCLUDING BUT NOT LIMITED TO DIRECT, INDIRECT, INCIDENTAL, PUNITIVE AND CONSEQUENTIAL DAMAGES. NO ORAL ADVICE OR WRITTEN INFORMATION GIVEN BY Bhavani Shankar Kodali SHALL CREATE A WARRANTY.


The information in capnography.com is intended to be used for review for personal study, credentialling exams and for entertainment. This makes no guarantee and gives no expectation to the reader, that information will result in successful completion of any credentialling exam for that reader.

Bhavani Shankar Kodali does not claim that the information supplied in the websites is sufficient to enable a reader to pass any credentialling exam or remain current in this field.

The information in this website is NOT intended to be used for clinical practice or patient care. There is also no intent to enter into any physician-patient relationship. Furthermore, the information is not to be used for purchasing stocks or any business dealing. Bhavani Shankar Kodali has made extensive efforts to ensure that the information is accurate and conform to the standards accepted at the time of preparation. However, constant changes in information resulting from continuing research and clinical experience, reasonable differences in opinions among authorities, unique aspects of individual clinical situations, and the possibility of human error in preparing such an extensive review require that the reader exercise individual judgment when making a clinical decision and, if necessary, consult and compare information from other sources. In particular, the reader is advised to check the product information provided by the capnograph manufacturer.

Under no circumstances will Bhavani Shankar Kodali of capnography.com be liable for any direct, indirect, special or consequential damages that result from technical, hardware or software failures of any kind; lost or unavailable network connections; or incomplete, garbled or delayed computer transmissions.

No Patient-Physician Relationship

By using this site, the user certifies that his or her intent is not to enter into any patient-physician relationship. As such, the user understands and accepts that this site does not substitute, augment or replace a patient-physician relationship.

Applicable Law

This site is created and controlled by Bhavani Shankar Kodali in the Commonwealth of Massachusetts. As such, the laws of the Commonwealth of Massachusetts will govern these disclaimers, terms and conditions, without giving effect to any principles of conflicts of laws. Bhavani Shankar Kodali reserves the right to make changes to this site and these disclaimers, terms and conditions at any time.