Anesthesia breathing circuits

ravititle

Table of contents

Introduction

Definition

Requirements of a Breathing System

Classification (Table 1)


Breathing systems without co2 absorption

Unidirectional Flow

Bidirectional Flow

Subclassification

Afferent reservoir system

Magill system

Enclosed afferent reservoir system

Efferent reservoir system

Bain system

Combined system

Breathing systems with co2 absorption

Circle system

Totally closed system

References

 

INTRODUCTION

Since the introduction of diethyl ether as an anaesthetic in 1846, the speciality of anaesthesia has come a long way. In the initial phase, the attention was mainly diverted to administering a single agent and apparatus were developed to suit the purpose. The reintroduction of nitrous oxide in 1868, and facility to store it in cylinders, created an interest in administering a combination of agents to anaesthetise patients. Any resemblance to a breathing system was developed by Barth (1907) using his valve with a nitrous oxide cylinder, a reservoir bag and Clover’s inhaler. Different positions of the lever in the valve allowed a range from complete rebreathing to completely breathing from the atmosphere. The development of Boyle’s machine (1917) and mastering of endotracheal intubation with a soft red rubber single lumen tube by Magill and Rowbothom were the forerunners for development of a simple anaesthetic delivery system by Magill, popularly known as the “Magill’s circuit”. Introduction of cyclopropane in 1929 and cuffed endotracheal tube in 1931; prompted Waters to develop the “to and fro” canister and use it for closed system anaesthesia with cyclopropane. In 1936, Brian Sword introduced the circle system. The Ayre’s T-piece was introduced in 1937, EMO inhaler in 1941 and Minnitt’s ‘gas and air’ apparatus with demand valve in 1949.

With the introduction of many breathing systems attempts were made to classify them in the 50’s and 60’s, but lack of a proper definition, lead to more confusion than clarity.

[Back]

DEFINITION:

A breathing system is defined as an assembly of components which connects the patient’s airway to the anaesthetic machine creating an artificial atmosphere, from and into which the patient breathes.

It primarily consists of

a) A fresh gas entry port/delivery tube through which the gases are delivered from the machine to the systems;

b) A port to connect it to the patient’s airway;

c) A reservoir for gas, in the form of a bag or a corrugated tube to meet the peak inspiratory flow requirements;

d) An expiratory port/valve through which the expired gas is vented to the atmosphere;

e) A carbon dioxide absorber if total rebreathing is to be allowed and

f) Corrugated tubes for connecting these components.

Flow directing valves may or may not be used.

[Back]

REQUIREMENTS OF A BREATHING SYSTEM:

The components when assembled should satisfy certain requirements, some essential and others desirable.

Essential:

The breathing system must

a) deliver the gases from the machine to the alveoli in the same concentration as set and in the shortest possible time;

b) effectively eliminate carbon-dioxide;

c) have minimal apparatus dead space; and

d) have low resistance.

Desirable:

The desirable requirements are

a) economy of fresh gas;

b) conservation of heat;

c) adequate humidification of inspired gas;

d) Light weight;

e) convenience during use;

f) efficiency during spontaneous as well as controlled ventilation (Efficiency is determined in terms of co2 elimination and fresh gas utilization);

g) adaptability for adults, children and mechanical ventilators;

h) provision to reduce theatre pollution.

[Back]


CLASSIFICATION OF BREATHING SYSTEMS:

One will realize the reason for the failure of the attempts at classification in the 50’s to 60’s, if this definition and requirements are taken into account. There are numerous classifications of breathing systems according to the whims and fancy of the person classifying. Many of them are irrelevant as they do not define a breathing system. Different authors classified the same system under different headings, adding to confusion1. McMohan in 1951 classified them as open, semiclosed and closed taking the level of rebreathing into account. It as follows:

Open
no rebreathing
Semiclosed
partial rebreathing
Closed total rebreathi


Dripps et al have classified them as Insufflation, Open, Semiopen, Semiclosed and Closed taking into account the presence or absence of Reservoir, Rebreathing, co2 absorption and Directional valves1. The ambiguity of the terminology used as open, semi open, semi closed and closed allowed inclusion of apparatus that are not breathing systems at all into the classification.

To overcome this problem Conway2 suggested that a functional classification be used and classified according to the method used for co2 elimination as:

1. Breathing systems with co2 absorber
2. Breathing systems without co2 absorber.

Miller.D.M.3 in 1988 widened the scope of this classification so as to include the enclosed afferent reservoir system.

A new breathing system called ‘The Maxima’4 has been designed by Miller in 1995 and to include it in the classification5, the enclosed afferent reservoir systems have been grouped under ‘displacement afferent reservoir’ systems.

Table 1. Classification of breathing systems



This classification also has a personal bias as the Humphrey ADE system is not included in the classification, even though he preferred to compare his system with that of Humphrey’s6. The classification suggested in table.1. is a partial modification of Miller’s3 classification.

 

BREATHING SYSTEMS WITHOUT co2 ABSORPTION.
BREATHING SYSTEMS WITH co2 ABSORPTION.

Unidirectional flow:

a) Non rebreathing systems.
B) Circle systems.

Unidirectional flow

Circle system with absorber.

Bi-directional flow:

a) Afferent reservoir systems.

Mapleson A

Mapleson B

Mapleson C

Lack’s system.

B) Enclosed afferent reservoir systems

Miller’s (1988)

c) Efferent reservoir systems

Mapleson D

Mapleson E

Mapleson F

Bain’s system

d) Combined systems

Humphrey ADE

Bi-directional flow

To and Fro system.

[Back]

BREATHING SYSTEMS WITHOUT co2 ABSORPTION

Unidirectional Flow

FIG 1. Nonrebreathing system. (Inset – Nonrebreathing Valve)

breath1


a) Nonrebreathing systems.

They use non rebreathing valves and there is no mixing of fresh gas and the expired gas.

 

Functional analysis:

When the patient takes a breath, or if the reservoir bag is squeezed, the inspiratory unidirectional valve opens and the gases flow into the patient’s lungs (Fig.1). The expiratory unidirectional valve closes the expiratory port during spontaneous breathing. The inspiratory unidirectional valve itself closes the expiratory port during controlled ventilation. At the start of expiration, the inspiratory unidirectional valve returns back to position and expiration takes place through the expiratory port, opening the expiratory valve.

The fresh gas flow (FGF) should be equal to the minute ventilation (MV) of the patient. These systems satisfy all four essential requirements, but are not very popular because of the following reasons:

1) Fresh gas flow has to be constantly adjusted and is not economical.

2) There is no humidification of inspired gas.

3) There is no conservation of heat.

4) They are not convenient as the bulk of the valve has to be positioned near the patient.

5) The valves can malfunction due to condensation of moisture and lead to complications.

b) Circle Systems:

These systems are designed with a co2 absorber as an essential component of the system. To use it without absorber is uneconomical as it needs a FGF more than the alveolar ventilation.

The effect of arrangements of various components in the effective elimination of co2 and fresh gas economy when used with high flows were analysed by Egar and Ethans7. Detailed discussion on this is beyond the purview of this review. However, some aspects of this is discussed under the section, Circle system with absorber.

 

[Back]

Bi-Directional Flow:

Systems with bi-directional flow are extensively used. These systems depend on the FGF for effective elimination of co2. Understanding these systems is most important as their functioning can be manipulated by changing parameters like Fresh gas flow, alveolar ventilation, apparatus dead space, etc. We will analyze these in detail.

Fresh Gas Supply:

Fresh gas flow (FGF) forms one of the essential requirements of a breathing system. If there is no FGF into the system, the patient will get suffocated. If the FGF is low, most systems do not eliminate carbon-dioxide effectively, and if there is an excess flow there is wastage of gas. So, it becomes imperative to specify optimum FGF for a breathing system for efficient functioning.

If the system has to deliver a set concentration in the shortest possible time to the alveoli, the FGF should be delivered as near the patient’s airway as possible.

Elimination Of Carbon-Dioxide:

The following may be taken as an example for better understanding of co2 elimination by the bi-directional flow systems. Normal production of carbon-dioxide in a 70 kg adult is 200 ml per minute and it is eliminated through the lungs. Normal end-tidal concentration of carbon-dioxide is 5%. Hence, for eliminating 200 ml of carbon-dioxide as a 5% gas mixture, the alveolar ventilation has to be:

200 x 100 = 4,000 ml.
5

This 4000 ml or 4 litres is the normal alveolar ventilation. Any breathing system connected to an adult’s airway should provide a minimum of 4 litres per minute of carbon-dioxide free gas to the alveoli for eliminating carbon-dioxide. If the alveolar ventilation becomes less than 4 litres per minute, it would lead to hypercarbia. If the alveoli are ventilated with 5 litres/minute of a gas containing 1% carbon-dioxide, or 8 litres/minute of a gas containing 2.5% carbon-dioxide, it could still eliminate 200 ml of carbon-dioxide per minute from the alveoli. It may be construed as 4 litres of co2 free gas and 1 litre of gas with 5% co2 in the first instant and as 4 litres of co2 free gas and 4 litres of gas with 5% co2 in the second instant. In effect, 4 litres of alveolar ventilation with co2 free gas is provided in both cases.

 

Apparatus Dead Space: It is the volume of the breathing system from the patient-end to the point up to which, to and fro movement of expired gas takes place.

FIG 2. Extent of dead space in various systems

breath2

In an afferent reservoir system with adequate FGF, the apparatus dead space extends up to the expiratory valve positioned near the patient (fig.2).

 

 

If the FG enters the system near the patient-end as in an efferent reservoir system, the dead space extends upto the point of FG entry. In systems where inspiratory and expiratory limbs are separate, it extends upto the point of bifurcation. The dynamic dead space will depend on the FGF and the alveolar ventilation. The dead space is minimal with optimal FGF. If the FGF is reduced below the optimal level, the dead space increases and the whole system will act as dead space if there is no FGF. Increasing the FGF above the optimum level will only lead to wastage of FG.

[Back]

Sub-Classification Of Bi-Directional Flow Systems:

Mapleson8 did a theoretical analysis of the fresh gas requirements of the semiclosed systems available at that time. It is only proper to refer to it as Mapleson systems as he gave a nomenclature as A, B, C, D and E for easy identification as per their construction. For better understanding of the functional analyses, they have been classified as:

1 Afferent reservoir system (ARS).
2 Enclosed afferent reservoir systems (EARS).
3 Efferent reservoir systems (ERS).
4 Combined systems.


The afferent limb is that part of the breathing system which delivers the fresh gas from the machine to the patient. If the reservoir is placed in this limb as in Mapleson A, B, C and Lack’s systems, they are called afferent reservoir systems (ARS).

The efferent limb is that part of the breathing system which carries expired gas from the patient and vents it to the atmosphere through the expiratory valve/port. If the reservoir is placed in this limb as in Mapleson D, E, F and Bain systems, they are called efferent reservoir systems (ERS).

Enclosed afferent reservoir system has been described by Miller and Miller.

[Back]

AFFERENT RESERVOIR (AR) SYSTEMS

The Mapleson A, B and C systems have the reservoir in the afferent limb, and do not have an efferent limb (Fig.3). Lack system has an afferent limb reservoir and an efferent limb through which the expired gas traverses before being vented into the atmosphere (Fig.6). This limb is coaxially placed inside the afferent limb.

FIG 3. Afferent reservoir systems (Mapleson’s A, B and C systems)

breath3

These AR systems work efficiently during spontaneous breathing provided the expiratory valve is separated from the reservoir bag and FGF by at least one tidal volume of the patient and apparatus dead space is minimal. They do not function efficiently during controlled ventilation. If the FGF is close to the expiratory valve as in Mapleson B & C, the system is inefficient both during spontaneous and controlled ventilation. The efficiency is determined in terms of carbon-dioxide elimination and FGF utilization.

 

Mapleson8 has analysed these bi-directional flow systems using mathematical calculations. He made a few basic assumptions while analyzing breathing systems. These are

(1) Gases move enblock. They maintain their identity as fresh gas, dead space gas and alveolar gas. There is no mixing of these gases.

(2) The reservoir bag continues to fill up, without offering any resistance till it is full.

(3) The expiratory valve opens as soon as the reservoir bag is full and the pressure inside the system goes above atmospheric pressure.

(4) The valve remains open throughout the expiratory phase without offering any resistance to gas flow and closes at the start of the next inspiration.

[Back]

Mapleson ‘A’/Magill’s system:

Functional analysis:

Spontaneous breathing: The system is filled with fresh gas before connecting to the patient. When the patient inspires, the fresh gas from the machine and the reservoir bag flows to the patient, and as a result the reservoir bag collapses (Fig.4a). During expiration, the FG continues to flow into the system and fill the reservoir bag. The expired gas, initial part of which is the dead space gas, pushes the FG from the corrugated tube into the reservoir bag and collects inside the corrugated tube (Fig.4b).

FIG 4.Functional analysis of Mapleson A (Magill system), Spontaneous breathing.

FGF fg Dead space gas ds Alveolar gas

breath7

 

As soon as the reservoir bag is full, the expiratory valve opens and the alveolar gas is vented into the atmosphere (Fig.4c). During the expiratory pause, alveolar gas that had come into the corrugated tube is also pushed out through the valve, depending on the FGF. The system is filled with only fresh gas and dead space gas at the start of the next inspiration when FGF is equal to the alveolar ventilation (Fig.4d). The entire alveolar gas and dead space gas is vented through the valve and some FG also escapes, if the FGF is higher than the minute ventilation. Some amount of alveolar gas will remain in the system and lead to rebreathing with a FGF less than the alveolar ventilation. This has been confirmed theoretically and experimentally by many investigators8,9. The system functions at maximum efficiency, when the FGF equals the alveolar ventilation and the dead space gas (which has not taken part in gas exchange) is allowed to be rebreathed and utilized for alveolar ventilation.

Controlled ventilation: To facilitate IPPV the expiratory valve has to be partly closed. During inspiration, the patient gets ventilated with FG and part of the FG is vented through the valve (Fig.5a) after sufficient pressure has developed to open the valve. During expiration, the FG from the machine flows into the reservoir bag and all the expired gas (i.e., dead space gas and alveolar gas) flows back into the corrugated tube till the system is full (Fig.5b). During the next inspiration the alveolar gas is pushed back into the alveoli followed by the FG. When sufficient pressure is developed, part of the expired gas and part of the FG escape through the valve (Fig.5c). This leads to considerable rebreathing, as well as excessive waste of fresh gas. Hence these systems are inefficient for controlled ventilation.

FIG 5. Mapleson A (Magill system), Controlled ventilation.

FGF fg Dead space gas ds Alveolar gas ag

breath11


 

Lack’s system:

This system functions like a Mapleson A system both during spontaneous and controlled ventilation. The only difference is that the expired gas instead of getting vented through the valve near the patient, is carried by an efferent tube placed coaxially and vented through the valve placed near the machine end (Fig.6). This facilitates easy scavenging of expired gas.

FIG 6. Lack’s system.

breath12

 

Mapleson B & C systems:

In order to reduce the rebreathing of alveolar gas and to improve the utilization of FG during controlled ventilation, the FG entry was shifted near the patient(Fig.3). This allows a complete mixing of FG and expired gas. The end result is that these systems are neither efficient during spontaneous nor during controlled ventilation.

[Back]

ENCLOSED AFFERENT RESERVOIR (EAR) SYSTEMS

This has been described by Miller & Miller10. The system consisted of a Mapleson A system enclosed within a non distensible structure (Fig.7a). It may also be constructed by enclosing the reservoir bag alone in a bottle and connecting the expiratory port to the bottle with a corrugated tube and a one way valve (Fig.7b). To the bottle is also attached a reservoir bag and a variable orifice for providing positive pressure ventilation.

FIG 7. Enclosed afferent reservoir system.

breath13

Functional analysis:

During spontaneous ventilation, the gas is vented from the system in a manner which is identical to the Mapleson A system. In this mode the variable orifice is kept widely open to allow free communication to the atmosphere. In controlled ventilation the reservoir bag ‘B’ is squeezed intermittently and the variable orifice is partly closed to allow building up of pressure in the bottle. The pressure thus developed (1) closes the expiratory valve, (2) squeezes the enclosed afferent reservoir and the patient gets ventilated. The expiration takes place in a manner similar to that described during spontaneous ventilation when the pressure is released in reservoir ‘B’,. Hence this system should function efficiently during spontaneous and controlled ventilation with a FGF equivalent to alveolar ventilation. The fresh gas requirement and the utilization of this system has been investigated by a group of investigators from Manchester11-13 and a group from Wales14 under the guidance of Mapleson. They have reported varying figures for utilization as 82%, 93% and 74% respectively11,12,14. The reasons for this lesser percentage of utilization have been quoted as faulty methodology for calculation15, resistance offered by the reservoir bag and tubing and early opening of the unidirectional valve during expiration14 etc. Though the fresh gas requirement is higher than the alveolar ventilation in this system as shown by the above studies, it is still more efficient than the Bain system for controlled ventilation.

[Back]

EFFERENT RESERVOIR (ER) SYSTEMS:

FIG 8. Efferent reservoir systems (Mapleson’s D, E and F)

breath14

FIG 9. Bain system


breath15




 

The Mapleson D, E, F and Bain systems have a 6 mm tube as the afferent limb that supplies the FG from the machine. The efferent limb is a wide-bore corrugated tube to which the reservoir bag is attached and the expiratory valve is positioned near the bag. In Mapleson E system, the corrugated tube itself acts as the reservoir (Fig.8). In Bain system, the afferent and efferent limbs are coaxially placed (Fig.9).All these ER systems are modifications of Ayre’s T-piece. This consists of a light metal tube 1 cm in diameter, 5 cm in length with a side arm (Fig.10). Used as such, it functions as a non-rebreathing system. Fresh gas enters the system through the side arm and the expired gas is vented into the atmosphere and there is no rebreathing. The dead space is minimal as it is only up to the point of FG entry and elimination of co2 is achieved by breathing into the atmosphere. FGF equal to peak inspiratory flow rate of the patient has to be used to prevent air dilution.

 

FIG 10. Ayre’s ‘T’ piece.

breath16

In an attempt to reduce FGF requirements, ER systems are constructed with reservoirs in the efferent limb. The functioning of all these systems are similar. These systems work efficiently and economically for controlled ventilation as long as the FG entry and the expiratory valve are separated by a volume equivalent to at least one tidal volume of the patient. They are not economical during spontaneous breathing.

 

[Back]

Bain system (Mapleson D)

Spontaneous respiration: The breathing system should be filled with FG before connecting to the patient. When the patient takes an inspiration, the FG from the machine, the reservoir bag and the corrugated tube flow to the patient (Fig.11a). During expiration, there is a continuous FGF into the system at the patient end. The expired gas gets continuously mixed with the FG as it flows back into the corrugated tube and the reservoir bag (Fig.11b). Once the system is full the excess gas is vented to the atmosphere through the valve situated at the end of the corrugated tube near the reservoir bag. During the expiratory pause the FG continues to flow and fill the proximal portion of the corrugated tube while the mixed gas is vented through the valve (Fig.11c). During the next inspiration, the patient breaths FG as well as the mixed gas from the corrugated tube (Fig.11d). Many factors influence the composition of the inspired mixture. They are FGF, respiratory rate, expiratory pause, tidal volume and co2 production in the body. Factors other than FGF cannot be manipulated in a spontaneously breathing patient. It has been mathematically calculated and clinically proved8,16 that the FGF should be atleast 1.5 to 2 times the patient’s minute ventilation in order to minimise rebreathing to acceptable levels.

FIG 11Functional analysis, Bain’s system (Mapleson D), Spontaneous breathing.

FGF fg Dead space gas ds Alveolar gas ag

breath20


Controlled ventilation: To facilitate intermittent positive pressure ventilation, the expiratory valve has to be partly closed so that it opens only after sufficient pressure has developed in the system. When the system is filled with fresh gas, the patient gets ventilated with the FGF from the machine, the corrugated tube and the reservoir bag (Fig.12a). During expiration, the expired gas continuously gets mixed with the fresh gas that is flowing into the system at the patient end. During the expiratory pause the FG continues to enter the system and pushes the mixed gas towards the reservoir (Fig.12B). When the next inspiration is initiated, the patient gets ventilated with the gas in the corrugated tube i.e., a mixture of FG, alveolar gas and dead space gas (Fig.12c). As the pressure in the system increases, the expiratory valve opens and the contents of the reservoir bag are discharged into the atmosphere.

 

FIG 12. Bain’s system, Controlled ventilation.

FGF fg Dead space gas ds Alveolar gas ag

breath24

Factors that influence the composition of gas mixture in the corrugated tube with which the patient gets ventilated are the same as for spontaneous respiration namely FGF, respiratory rate, tidal volume and pattern of ventilation. The only difference is that these parameters can be totally controlled by the anaesthesiologist and do not depend on the patient. Using a low respiratory rate with a long expiratory pause and a high tidal volume, most of the FG could be utilized for alveolar ventilation without wastage.

 

FIG 13. Relation between alveolar ventilation and FGF

breath25

Analyzing the performance of these systems during controlled ventilation, two relationships have become evident. 1) When FGF is very high the Paco2 becomes ventilation dependent (as during spontaneous respiration). 2) When the minute volume exceeds the FGF substantially, the Paco2 is dependent on the FGF17. Combining these influences a graph can be constructed as shown in Fig.13. An infinite number of combinations of FGF and minute ventilation can be chosen to achieve a desired Paco2. One can use a high FGF and a normal minute volume of 70 ml/kg to achieve a normal Paco2 of 40 mm Hg. This is uneconomical and leads to low humidity and heat loss. Alternately, a FGF equivalent to the predicted minute volume i.e., 70 ml/kg can be chosen and the patient ventilated with at least twice the predicted minute volume i.e. 140 ml/kg. Here a deliberate controlled rebreathing is allowed in order to maintain normal Paco2 along with high humidity, less heat loss and greater economy of fresh gas. Combinations between these two extremes can also be used. It is important to remember that using a low FGF with normal minute ventilation, can lead to hypercarbia; a moderate FGF and hyperventilation, can lead to hypocarbia.

 

[Back]

COMBINED SYSTEMS

To over come the difficulties of changing the breathing systems for different modes of ventilation, Humphrey designed a system called Humphrey ADE18, with two reservoirs, one in the afferent limb and the other in the efferent limb. While in use, only one reservoir will be in operation and the system can be changed from ARS to ERS by changing the position of a lever. It can be used for adults as well as children. The functional analysis is the same as Mapleson A in ARS mode and as Bain in ERS mode. It is not yet widely used.

[Back]

BREATHING SYSTEMS WITH co2 ABSORPTION

Systems so far described have relied on FGF for effective elimination of co2. Any desire to economize on FGF by allowing a total rebreathing, should be accompanied by removal of the expired co2 by chemical absorption using sodalime or baralyme. The systems designed for these purpose are again classified as:

Unidirectional flow.

-Circle system.

Bi-directional flow.

-To and fro system.

Circle System

The essential components of the circle system are, (1) a sodalime canister, (2) Two unidirectional valves, (3) Fresh gas entry, (4) Y-piece to connect to the patient, (5) Reservoir bag (6) a relief valve and (7) low resistance interconnecting tubing. The arrangement of the components is shown in fig.14.

FIG 14. Components of circle system

breath26

For efficient functioning of the system the following criteria should fulfilled. (1) There should be two unidirectional valves on either side of the reservoir bag, (2) Relief valve should be positioned in the expiratory limb only, (3) The FGF should enter the system proximal to the inspiratory unidirectional valve.

 

Functional analysis:

During inspiration the FG along with the co2 free gas in the reservoir bag flow through the inspiratory limb and inspiratory unidirectional valve to the patient. No flow takes place in the expiratory limb as the expiratory unidirectional valve is closed by back pressure transmitted to the valve. During expiration the inspiratory unidirectional valve closes and the expired gas flows through the expiratory unidirectional valve in the expiratory limb to the sodalime canister and to the reservoir bag. The co2 is absorbed in the canister. The FGF from the machine continues to fill the reservoir bag. When the reservoir is full the relief valve opens and the excess gas is vented to atmosphere. By selecting a suitable position for the relief valve, the expired gas can be selectively vented when the FGF is more than the alveolar ventilation. To facilitate controlled ventilation the relief valve has to be partly closed and the excess gas is vented during inspiration. The gas flow pattern is similar to that described above.

The advantages and disadvantages of the various arrangements of the components were analyzed by Eger and Ethans7. The relative positions of the components of the circle system are of particular importance to the functioning of the system only when the FGF is high, the gas components of the system unmixed and co2 absorber not used. When the FGF is reduced below the alveolar ventilation, the co2 absorber is a must as the gas in the system become more uniformly mixed, and the relative position of the system’s components become less important

[Back]

Totally closed system:

The systems with co2 absorption can be used in a completely closed mode. After a period of approximately 10-20 minutes breathing with high inflow of fresh gas for denitrogenation, the expiratory valve is closed. The FGF is then adjusted to meet only the patients basal oxygen requirements together with anaesthetic. A number of advantages have been demonstrated for totally closed systems.

A) Economy: The FGF could be reduced to as low as 250 - 500 ml of oxygen. The consumption of Halothane/Isoflurane has been found to be around 3.5 ml/hour19.

b) Humidification: In the completely closed system, once the equilibrium has been established, the inspired gas will be fully saturated with water vapour20.

C) Reduction of heat loss: In addition to conserving water the totally closed system will also conserve heat. The co2 absorption is an exothermic reaction and the system may actively assist in maintaining body temperature.

D) Reduction in atmospheric pollution: Once the expiratory valve has been closed, no anaesthetic escapes, except for the small percutaneous loss from the patient.

E) Control of anaesthesia: It is possible to compute the time course of uptake of anaesthetic in a patient of known size and add the appropriate quantity of the anaesthetic to the circuit at a rate decreasing in a manner calculated to maintain a constant alveolar concentration21. In practice an alveolar concentration of about 1.3 x MAC is found to be suitable.

The technique has several potential disadvantages.

i) A greater knowledge of uptake and distribution is required to master closed circuit anaesthesia.

ii) Inability to alter any concentration quickly.

iii) Real danger of hypercapnia may result from, a) an inactive absorber, B) incompetent unidirectional valves and c) incorrect use of absorber bypass.

Bi-directional flow systems

The Waters to and fro system is valveless and conveniently portable. It has been widely used in the past and now is only of historical importance. The reader may refer to any standard text book for furtherdetails.


[Back]

References:

1. Dorsch, J.A., Dorsch,S.E. Breathing systems II. In understanding anaesthesia equipment. 2nd ed. 1984. Williams & Wilkins.

2. Conway, C.M.: Anaesthetic breathing systems. British Journal of Anaesthesia. 1985, 57, 649-57.

3. Miller, D.M.: Breathing systems for use in anaesthesia. Evaluation using a physical lung model and classification. British Journal of Anaesthesia. 1988, 60, 555-64.

4. Miller DM. An enclosed efferent afferent reservoir system: The Maxima. Anaesthesia and intensive care 1995; 23:284-91.

5. Miller DM. Breathing systems reclassified. Anaesthesia and intensive care 1995; 23: 281-83.

6. Miller DM, Palm A. Comparison in spontaneous ventilation of the Maxima with the Humphrey ADE breathing system and between four methods for detecting rebreathing. Anaesthesia and intensive care 1995; 23: 296-01.

7. Eager EI., Ethans, C.T.: The effects of inflow, overflow and valve placement on economy of circle system. Anesthesiology, 1968, 29, 93-100.

8. Mapleson WW. The elimination of rebreathing in various semiclosed anaesthetic systems. British journal of Anaesthesia; 1954;26: 323-32.

9. White, DC, Calkins, J. Anaesthetic machine and breasting apparatus. In: Nunn, Utting and Brown eds. General anaesthesia. 5th ed.. Butterworths, London 1989:428-56

10. Miller, D.M., Miller, J.C.: Enclosed afferent reservoir breathing systems. Description and clinical evaluation. British Journal of Anaesthesia. 1988, 60, 469-75.

11. Droppert PM, Meakin G, Beatty PCW, Mortimer AJ, Healy TEJ. Efficiency of a new afferent reservoir breathing system during controlled ventilation. British Journal of Anaesthesia 1991; 66: 638-42.

12. Meakin G, Jennings AD, Beatty PCW, Healy TEJ. Fresh gas requirements of an enclosed afferent reservoir breathing system during controlled ventilation in children. British Journal of Anaesthesia 1992; 68: 43-47.

13. Beatty PCW, Meakin G, Healy TEJ. Fractional delivery of fresh gas: a new index of the efficiency of semiclosed breathing systems. British Journal of Anaesthesia 1992; 68: 474-77.

14. Tham EJ, Davies R, Slade JM, Mapleson WW. Efficiency of breathing systems A and D in the Carden Ventmasta ventilator. British Journal of Anaesthesia 1993; 71: 741-46.

15. Ravishankar M, Chatterjee S. Fractional utilisation of fresh gas by breathing systems without carbon dioxide absorption. British Journal of Anaesthesia 1993; 71: 706-07

16. Ward CS. In: Anaesthetic equipment. Physical principles and maintenance; W.B.Saunders, London; 2nd ed. 1985.

17. Rose DK, Froese AB. The regulation of Paco2 during controlled ventilation of children with a T-piece. Canadian Anaesthetists Society Journal 1979; 26: 104-13.

18. Humphrey D. A new anaesthetic breathing system combining Mapleson A,D and E principles. Anaesthesia 1983; 38: 361-72

19. Baum JA, Aitkenhead AR. Low flow anaesthesia. Anaesthesia 1995; 50(supplement):37-44.

20. Kleemann, P.P. Humidity of anaesthetic gases with respect to low flow anaesthesia. Anaesthesia and Intensive Care, 1994, 22 (4), 396-408.

21. Lowe HJ, Ernst EA. The quantitative practice of anaesthesia, Use of closed circuit. Baltimore. Williams & wilkins, 1981.

 

Anesthesia Breathing Systems

 

Anesthesia Breathing Systems

M Ravi Shankar MD  and Bhavani Shankar Kodali MD

 

M Ravi Shankar MD, Formerly Professor and Chair of Anesthesia,  JIPMER, India, and presently Presently Professor and Chair of Anesthesiology, Mahatma Gandhi Medical College and Research Institute

Pondicherry, India 

 

 

For a comprehensive understanding of 'Anesthesia breathing circuits' click here for a review. It opens a new window, close the window to return to this section again. 


The following section contains pertinent features of anesthesia circuits and their functional analysis for easy understanding facilitated by animations.

DEFINITION

A breathing system is defined as an assembly of components which connects the patient’s airway to the anaesthetic machine creating an artificial atmosphere, from and into which the patient breathes

It primarily consists of
a) A fresh gas entry port/delivery tube through which the gases are delivered from the machine to the systems;

b) A port to connect it to the patient’s airway;

c) A reservoir for gas, in the form of a bag or a corrugated tube to meet the peak inspiratory flow requirements;

d) An expiratory port/valve through which the expired gas is vented to the atmosphere;

e) A carbon dioxide absorber if total rebreathing is to be allowed and

f) Corrugated tubes for connecting these components.

Flow directing valves may or may not be used.

Classification

BREATHING SYSTEMS WITHOUT CO2 ABSORPTION
BREATHING SYSTEMS WITH CO2 ABSORPTION.
Unidirectional flow:

a) Non rebreathing systems.
B) Circle systems.

Unidirectional flow


Circle system with absorber.

Bi-directional flow:


a) Afferent reservoir systems.

Mapleson A

Mapleson B

Mapleson C

Lack’s system.

B) Enclosed afferent reservoir systems

Miller’s (1988)

c) Efferent reservoir systems

Mapleson D

Mapleson E

Mapleson F

Bain’s system

d) Combined systems

Humphrey ADE
Bi-directional flow

To and Fro system.

<image>

Mapleson circuits

In 1954 Mapleson described and analyzed five different semiclosed anesthetic systems, and they are classically referred to as Mapleson systems and designated A to E. In 19 , Willis et al added the F system to the five original systems. The table below shows appearance and description of each system. For functional analysis of each of these systems, click on the hyperlink to open a new window. Close the window to return back to this section.

There is mixing of these two in rebreathing systems

Type Appearance Description
Mapleson A
mpastatic
Expiratory valve close to face mask, separated by corrugated tube ( ) from reservoir bag and supply of fresh gases
Mapleson B
mpbstatic
Expiratory valve and supply of fresh gases close to face mask, separated by corrugated tube ( ) from reservoir bag
Mapleson C
mpcstatic
Expiratory valve, supply of fresh gases and reservoir bag , all close to face mask (No corrugated tube ( )
Mapleson D mpdstatic Supply of fresh gases close to face mask, separated by corrugated tube ( ) from reservoir bag and expiratory valve
Mapleson E mpestatic Supply of fresh gases close to face mask. Open length of corrugated tube ( ). No reservoir bag or expiratory valve.
Mapleson F mpfstatic Supply of fresh gases close to face mask. Corrugated tube ( ) present. . Reservoir bag with expiratory port, but no expiratory valve.

Mapelson E with reservoir bag.

Relative efficiency of rebreathing among various Mapleson circuits

Spontaneous ventilation A>DFE>CB
Controlled ventilation DFE>BC>

Mapleson A is most efficient during spontaneous ventilation. It is worst for controlled ventilation.

Mapleson D is most efficient during controlled ventilation

Circle system

Bain system

This is a modified Mapleson D system, with fresh gas flow delivery tube running inside of the expiratory limb
bain

 

Non rebreathing system

This is a unidirectional non rebreathing system with out CO2 absorption. A non rebreathing valve is located near the patient end.
nrb

For other circuits refer to the monogram

2003-2004


ASA abstracts 2003-2004

   Record 1       Year: 2003       Abstract Number: A-564       Score: 100
Title: Evaluation for New Main-Stream Capnography with Sedativenon-Intubated Volunteers in Comparison with Side-Stream Capnography
Authors(s): Nobutada Morioka, M.D., Shinji Yamamori, , Mitsuhiro Narushima, M.D., Chisato Nozaki, M.D., Makoto Ozaki, M.D., Ph.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 2       Year: 2004       Abstract Number: A-1503       Score: 64
Title: Usefulness of Oral and Nasal Capnography in the Pediatric Recovery Room
Authors(s): Florence Babre, M.D., Maryline Bordes, M.D., Francois Semjen, M.D., Anne-Marie Cros, M.D.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 3       Year: 2004       Abstract Number: A-510       Score: 35
Title: Aspiration Capnography Guides Intubation Accurately
Authors(s): Glen Atlas, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

Record 1       Year: 2004       Abstract Number: A-1556       Score: 100
Title: Arterial to End-Tidal Carbon Dioxide Difference during Thoracoscopic Surgery
Authors(s): Venkatesh Srinivasa, M.D., Bhavani-Shankar Kodali, M.D., Teresa Bean, M.D., Philip M. Hartigan, M.D.
Topic: RESPIRATION

View Files:    • HTML   |   Printable Version

   Record 2       Year: 2003       Abstract Number: A-524       Score: 100
Title: Recognizing an Increased Arterial to End-Tidal Carbon Dioxide Gradient (PaCO 2 - EtCO 2 ) Due to Equipment Failure: A Full Human
Authors(s): W. Bosseau Murray, MB, ChB, Terry T. Durbin, D.D.S., Jody L. Henry, B.S.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version
Record 1       Year: 2003       Abstract Number: A-564       Score: 100
Title: Evaluation for New Main-Stream Capnography with Sedativenon-Intubated Volunteers in Comparison with Side-Stream Capnography
Authors(s): Nobutada Morioka, M.D., Shinji Yamamori, , Mitsuhiro Narushima, M.D., Chisato Nozaki, M.D., Makoto Ozaki, M.D., Ph.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 2       Year: 2004       Abstract Number: A-1503       Score: 64
Title: Usefulness of Oral and Nasal Capnography in the Pediatric Recovery Room
Authors(s): Florence Babre, M.D., Maryline Bordes, M.D., Francois Semjen, M.D., Anne-Marie Cros, M.D.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 3       Year: 2004       Abstract Number: A-1559       Score: 35
Title: Assessment of Gas Exchange in Spontaneously Breathing Patients
Authors(s): Hartmut Gehring, M.D., Norman Stein, Holger Matz, Ph.D., Ewald Konecny, Ph.D.
Topic: RESPIRATION

View Files:    • HTML   |   Printable Version

   Record 4       Year: 2004       Abstract Number: A-1438       Score: 35
Title: A Third-Generation Acoustic Reflectometer for Pediatric Intubations
Authors(s): Gligor V. Gucev, M.D., Shlomo Elspas, M.D., David Raphael, M.D.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 5       Year: 2004       Abstract Number: A-510       Score: 35
Title: Aspiration Capnography Guides Intubation Accurately
Authors(s): Glen Atlas, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 6       Year: 2003       Abstract Number: A-395       Score: 35
Title: Capnometry in Non-Intubated Emergency Patients Improves Pre-Hospital Monitoring Quality Compared to Pulse Oxymetry
Authors(s): Alma Joldzo, B.S., Petra Bertalanffy, M.D., Burkhard Gustorff, M.D., Alexander Kober, M.D., Klaus Hoerauf, M.D., Ph.D.
Topic: CRITICAL CARE

View Files:    • HTML   |   Printable Version

   Record 7       Year: 2003       Abstract Number: A-1503       Score: 28
Title: Facilitation of Blind Intubation Via the LMA Fastrach® Using the Self-Inflating Bulb
Authors(s): Xuan Au-Truong, M.D., Xiaoyuan Xie, M.D., M. Ramez Salem, M.D., Ninos J. Joseph, B.S.
Topic: RESPIRATION

View Files:    • HTML   |   Printable Version

   Record 8       Year: 2004       Abstract Number: A-1494       Score: 21
Title: Children Sedated with Chloral Hydrate during Spine MRI Scans Have Increases in Their Core Body Temperature
Authors(s): Yvon F. Bryan, M.D., David B. Glick, M.D., Martin Szafran, B.S., Thomas Templeton, M.D., Avery Tung, M.D.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 9       Year: 2004       Abstract Number: A-1433       Score: 21
Title: Assessing the Knowledge of American Heart Association Cardiopulmonary Resuscitation Guidelines in Pediatric Anesthesiologists: A
Authors(s): Donald H. Shaffner, M.D., Eugenie S. Heitmiller, M.D., Lynne G. Maxwell, M.D., Elizabeth A. Hunt, M.D., M.P.H., Myron Yaster, M.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 10       Year: 2004       Abstract Number: A-983       Score: 21
Title: Predict Postoperative Surgical Pain by Preoperative Pressure Pain Measurement
Authors(s): Yu-Chun Hung, M.D., Chi Hang Kuok, M.D., Chien-Chuan Chen, M.D., Yung-Wei Hsu, M.D.
Topic: LOCAL ANESTHESIA AND PAIN

View Files:    • HTML   |   Printable Version
Record 11       Year: 2004       Abstract Number: A-603       Score: 21
Title: Detection of Esophageal Intubation Using Portable Ultrasonography (SonoSite 180 PLUS)
Authors(s): Hiroshi Makino, M.D., Takasumi Katoh, M.D., Shigehito Sato, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 12       Year: 2004       Abstract Number: A-517       Score: 21
Title: Comparison of the Esophageal-Tracheal Combitube and EasyTube in Paralyzed Anesthetized Adult Patients
Authors(s): Luis A. Gaitini, M.D., Sonia J. Vaida, M.D., Mostafa Somri, M.D., Boris Yanovski, M.D., Riad Toame, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 13       Year: 2004       Abstract Number: A-601       Score: 21
Title: Handheld Tesla Meter Confirmation of Tracheal Tube Placement
Authors(s): Takasumi Katoh, M.D., Hiroshi Makino, M.D., Shigehito Sato, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 14       Year: 2004       Abstract Number: A-575       Score: 21
Title: The Acoustic Sensor for Monitoring Ventilation of Separate Lungs
Authors(s): Shai M.R. Tejman-Yarden, M.D., Alex Zlotnik, M.D., Arnon Cohen, Ph.D., Gabriel M. Gurman, M.D.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 15       Year: 2004       Abstract Number: A-122       Score: 21
Title: The Quality of Recovery after Induction and Maintenance Anesthesia with Remifentanil and Desflurane or Sevoflurane
Authors(s): Raoul Muchada, M.D., Génvieve Lucchesi, M.D., Dominique Brunel, M.D.
Topic: ANESTHETIC ACTION AND BIOCHEMISTRY

View Files:    • HTML   |   Printable Version

   Record 16       Year: 2003       Abstract Number: A-1416       Score: 21
Title: Sedated and Not Cool: Infants and Toddlers Do Not Become Hypothermic during MRI When Sedated with Chloral Hydrate
Authors(s): Yvon Bryan, M.D., Thomas Templeton, M.D., Avery Tung, M.D.
Topic: PEDIATRIC ANESTHESIA

View Files:    • HTML   |   Printable Version

   Record 17       Year: 2003       Abstract Number: A-1256       Score: 21
Title: Prediction of Difficult Laryngoscopy in Obese Patients by Ultrasound Quantification of Anterior Neck Soft Tissue
Authors(s): Peter Szmuk, M.D., Sergio Susmallian, M.D., Gabriella Gewürtz, M.D., Daniel Sessler, M.D., Tiberiu Ezri, M.D.
Topic: PATIENT SAFETY, PRACTICE MANAGEMENT, HISTORY, and EDUCATION

View Files:    • HTML   |   Printable Version

   Record 18       Year: 2003       Abstract Number: A-524       Score: 21
Title: Recognizing an Increased Arterial to End-Tidal Carbon Dioxide Gradient (PaCO 2 - EtCO 2 ) Due to Equipment Failure: A Full Human
Authors(s): W. Bosseau Murray, MB, ChB, Terry T. Durbin, D.D.S., Jody L. Henry, B.S.
Topic: EQUIPMENT, MONITORING, and ENGINEERING TECHNOLOGY

View Files:    • HTML   |   Printable Version

   Record 19       Year: 2003       Abstract Number: A-280       Score: 21
Title: A Comparison of Sevoflurane, Target-Controlled Infusion Propofol, Anesthesia in Patients Undergoing Elective Brain Tumor Surgery
Authors(s): Benoit Rossignol, M.D., 1, Gildas Gueret, M.D., 2, Guy Le Gall, M.D., 3, Charles C. Arvieux, M.D., Ph.D., 4.
Topic: CLINICAL NEUROSCIENCES

View Files:    • HTML   |   Printable Version

Record 4       Year: 2004       Abstract Number: A-1557       Score: 42
Title: Partial Pressure End-Tidal CO 2 (PetCO 2 ) Accurately Predicts PaCO 2 for a Large Majority of Patients with Acute Respiratory Di
Authors(s): Michael J. Banner, Ph.D., Vic Brennan, Ph.D., Andrea Gabrielli, M.D.
Topic: RESPIRATION

View Files:    • HTML   |   Printable Version

   Record 5       Year: 2004       Abstract Number: A-1554       Score: 42
Title: A New Approach To Evaluate Early Postoperative Respiratory Compromise in Morbidly Obese Patients
Authors(s): John L. Chow, M.D., M.S., Deborah M. Williams, B.A., Hendrikus J. Lemmens, M.D., Ph.D., William Feaster, M.D., M.B.A., John G. B
Topic: RESPIRATION

In literature

New applications of capnography


Capnography versus bronchoscope during percutaneous tracheostomy

A prospective randomized controlled trial of capnography vs. bronchoscopy for Blue Rhino percutaneous tracheostomy.

Mallick A, Venkatanath D, Elliot SC, Hollins T, Nanda Kumar CG. A prospective randomized controlled trial of capnography vs. bronchoscopy for Blue Rhino percutaneous tracheostomy. Anaesthesia. 2003 Sep; 58(9): 864-8.

A crucial step for successful percutaneous tracheostomy is the introduction of the needle and guide wire into the trachea. Capnography has recently been proposed as one way to confirm tracheal needle placement. In this randomized controlled study, the authors used capnography in 26 patients and bronchoscopy in 29 patients to confirm needle placement for percutaneous tracheostomy using Blue Rhino kit. The operating times and the incidence of peri-operative complications were similar for both groups. Capnography proved to be as effective as bronchoscopy in confirming correct needle placement.

Capnography for feeding tube insertion


Pulmonary complications of feeding tubes: a new technique of insertion and monitoring malposition.

D'Souza CR, Kilam SA, D'Souza U, Janzen EP, Sipos RA. Can J Surg. 1994 Oct; 37(5): 404-8.

The study consisted of thirteen anesthetized adult patients and 7 awake subjects scheduled to undergo elective surgery. Airway sampling for carbon dioxide with capnography was performed in 13 anesthetized adults with the tip of the feeding tube in the pharynx, in the esophagus and in the trachea, and airway sampling for carbon dioxide from the pharynx and esophagus in 7 awake subjects during introduction of the feeding tube. Fluoroscopic monitoring of the position of the tip of the feeding tube was performed during introduction in two patients and two volunteers. In all patients, with the tube either in the trachea or pharynx, a normal capnogram was displayed. When the tube was introduced into the esophagus no capnogram curve was seen, indicating the absence of carbon dioxide. With the subject lying down during introduction, the weighted tube followed the posterior pharyngeal wall to the upper esophageal sphincter.

CONCLUSION: Positioning of the patient lying down with the head flexed and capnographic measurement of carbon dioxide levels from the tip of the feeding tube during insertion is a safe, accurate and cost-effective method for the introduction of feeding tubes.

Capnography confirms correct feeding tube placement in intensive care unit patients.

Kindopp AS, Drover JW, Heyland DK. Can J Anaesth. 2001

The authors tested the accuracy and potential time savings of capnography as compared with a two-step radiographic method in placing feeding tubes in critically ill patients. One hundred feeding tube placements were studied in a tertiary care intensive care unit. All placements utilized a two-step radiographic method, but capnography was added to the procedure. The procedure was then completed or abandoned depending on radiographic interpretation. Radiography showed 11 feeding tubes projecting within the tracheobronchial tree. In all 11 of these placements, the capnography unit displayed a normal capnogram. Radiography revealed 86 tube placements in the midesophageal region. In all 86 of these placements, capnography displayed a "purging warning". In three placements, radiography indicated that the tube was coiled in the oropharynx. In these cases, the capnograph displayed one "no purging/no capnogram" result, and two "purging" warnings. If using capnography alone, an average of 72.5 min would be required to complete a feeding tube placement (which includes time for requisite "pre-feed radiograph"). The two-step radiological approach took an average of 169.4 min, a difference of 96.9 min (P <0.0001) between the two methods.

CONCLUSIONS: Capnography accurately identified all intratracheal feeding tube placements in this study. This study also shows that the use of capnography would significantly shorten the time needed for tube placement compared with a two-step radiologic method. The authors conclude that capnography should be considered for routine use when placing feeding tubes since it adds little time to the procedure and may improve patient safety.



Report on the development of a procedure to prevent placement of feeding tubes into the lungs using end-tidal co2 measurements.


Burns SM, Carpenter R, Truwit JD. Crit Care Med. 2001 May; 29(5): 936-9.


: To determine the accuracy of a technique using capnography to prevent inadvertent placement of small-bore feeding tubes and Salem sump tubes into the lungs. : A total of 25 ventilated adult MICU patients were studied-5 in phase 1 and 20 in phase 2. Phase 1 tested the ability of the end-tidal co2 (ETco2) monitor to detect flow (and thus accurately detect co2) through small-bore feeding tubes. A small-bore feeding tube, with stylet in place, was placed 5 cm through the top of the tracheostomy tube ventilator adapter in five consecutive patients. The distal end of the feeding tube was attached to the ETco2 monitor. The ETco2 level and waveform were assessed and recorded. Because co2 waveforms were successfully detected, a convenience sample of 20 adult MICU patients who were having feeding tubes placed (13 Salem sump tubes, 7 small-bore feeding tubes) was then studied. The technique consisted of attaching the ETco2 monitor to the tubes and observing the ETco2 waveform throughout placement. Of the seven small-bore feeding tubes tested, all were successfully placed on initial insertion. Placement was confirmed by absence of an ETco2 waveform and by radiograph. Of the 13 Salem sump tubes, 9 were placed successfully on first attempt and confirmed by absence of co2 and by air bolus and aspiration of stomach contents. ETco2 waveforms were detected with insertion of four of the Salem sump tubes; the tubes were immediately withdrawn, and placement was reattempted until successful.

CONCLUSIONS: The technique described is a simple, cost-effective method of assuring accurate gastric tube placement in critically ill patients.