Functional analysis |
|||||||
|
|
|||||||
|
FGF should be equal to the minute ventilation of the patient |
Functional
analysis:
When the patient takes a breath, or if the reservoir bag is squeezed, the
inspiratory unidirectional valve opens and the gases flow into the patient’s
lungs(Fig.1). The expiratory
unidirectional valve closes the expiratory port during spontaneous breathing. The inspiratory unidirectional valve itself closes the
expiratory port during controlled ventilation.
At the start of expiration, the inspiratory unidirectional valve returns
back to position and expiration takes place through the expiratory port, opening
the expiratory valve.
The
fresh gas flow (FGF) should be equal to the minute ventilation (MV)
of the patient. These systems satisfy all four essential requirements, but are
not very popular because of the following reasons:
1)
Fresh gas flow has to be constantly adjusted and is not economical.
2)
There is no humidification of inspired gas.
3)
There is no conservation of heat.
4)
They are not convenient as the bulk of the valve has to be positioned near the
patient.
5)
The valves can malfunction due to condensation of moisture and lead to
complications.