Functional analysis |
|||||||
|
|
|||||||
|
Effect of FGF on capnogram |
|||||||
|
|
|
Rebreathing capnograms are due to |
| Exhausted soda lime |
![]() ![]() |
| Inspiratory valve malfunction |
![]() |
| Expiratory valve malfunction |
|
|
The
essential components of the circle system are, (1) a sodalime canister, (2) Two
unidirectional valves, (3) Fresh gas entry, (4) Y-piece to connect to the
patient, (5) Reservoir bag (6) a relief valve and (7) low resistance
interconnecting tubing. The arrangement of the components is shown in figure above.
For efficient functioning
of the system the following criteria should fulfilled.
(1) There should be two unidirectional valves on either side of the
reservoir bag, (2) Relief valve should be positioned in the expiratory limb
only, (3) The FGF should enter the system proximal to the inspiratory
unidirectional valve.
Functional
analysis: During inspiration the FG along with the CO2 free
gas in the reservoir bag flow through the inspiratory limb and inspiratory
unidirectional valve to the patient. No
flow takes place in the expiratory limb as the expiratory unidirectional valve
is closed by back pressure transmitted to the valve. During expiration the inspiratory unidirectional valve closes
and the expired gas flows through the expiratory unidirectional valve in the
expiratory limb to the sodalime canister and to the reservoir bag.
The CO2 is absorbed in the canister.
The FGF from the machine continues to fill the reservoir bag.
When the reservoir is full the relief valve opens and the excess gas is
vented to atmosphere. By selecting
a suitable position for the relief valve, the expired gas can be selectively
vented when the FGF is more than the alveolar ventilation. To facilitate
controlled ventilation the relief valve has to be partly closed and the excess
gas is vented during inspiration. The gas flow pattern is similar to that
described above.
The
advantages and disadvantages of the various arrangements of the components were
analyzed by Eger and Ethans1. The
relative positions of the components of the circle system are of particular
importance to the functioning of the system only when the FGF is high, the gas
components of the system unmixed and CO2 absorber not used.
When the FGF is reduced below the alveolar ventilation, the CO2
absorber is a must as the gas in the system become more uniformly mixed, and the
relative position of the system’s components become less important
Totally
closed system:
The
systems with CO2 absorption can be used in a completely closed mode.
After a period of approximately 10-20 minutes breathing with high inflow
of fresh gas for denitrogenation, the expiratory valve is closed.
The FGF is then adjusted to meet only the patients basal oxygen
requirements together with anaesthetic. A
number of advantages have been demonstrated for totally closed systems.
A)
Economy: The FGF could be reduced
to as low as 250 - 500 ml of oxygen. The
consumption of Halothane/Isoflurane has been found to be around 3.5 ml/hour19.
b)
Humidification: In the completely
closed system, once the equilibrium has been
established, the inspired gas will be fully saturated with water vapour20.
C)
Reduction of heat loss: In addition
to conserving water the totally closed system will also conserve heat. The CO2
absorption is an exothermic reaction and the system may actively assist in
maintaining body temperature.
D)
Reduction in atmospheric pollution: Once
the expiratory valve has been closed, no anaesthetic escapes, except for the
small percutaneous loss from the patient.
E)
Control of anaesthesia: It is
possible to compute the time course of uptake of anaesthetic in a patient of
known size and add the appropriate quantity of the anaesthetic to the circuit at
a rate decreasing in a manner calculated to maintain a constant alveolar
concentration21. In
practice an alveolar concentration of about 1.3 x MAC is found to be suitable.
The
technique has several potential disadvantages.
i)
A greater knowledge of uptake and distribution is required to master closed
circuit anaesthesia.
ii)
Inability to alter any concentration quickly.
iii)
Real danger of hypercapnia may result from, a) an inactive absorber, B)
incompetent unidirectional valves and c) incorrect use of absorber bypass.
1.
Eager EI., Ethans, C.T.: The
effects of inflow, overflow and valve placement
on economy of circle system. Anesthesiology,
1968, 29, 93-100.
2.
Baum JA, Aitkenhead AR. Low flow anaesthesia. Anaesthesia 1995;
50(supplement):37-44.
3.
Kleemann, P.P. Humidity of
anaesthetic gases with respect to low flow anaesthesia.
Anaesthesia and Intensive Care, 1994, 22 (4), 396-408.