|
ANAESTHETIC BREATHING SYSTEMS |
|
DR M Ravi Shankar Professor and chair of Anesthesia JIPMER, Pondicheryy, India
|
Introduction
Since
the introduction of diethyl ether as an anaesthetic in 1846, the speciality of
anaesthesia has come a long way. In the initial phase, the attention was mainly
diverted to administering a single agent and apparatus were developed to suit
the purpose. The reintroduction of nitrous oxide in 1868, and facility to store
it in cylinders, created an interest in administering a combination of agents to
anaesthetise patients. Any resemblance to a breathing system was developed by
Barth (1907) using his valve with a nitrous oxide cylinder, a reservoir bag and
Clover’s inhaler. Different positions of the lever in the valve allowed a
range from complete rebreathing to completely breathing from the atmosphere. The
development of Boyle’s machine (1917) and mastering of endotracheal intubation
with a soft red rubber single lumen tube by Magill and Rowbothom were the
forerunners for development of a simple anaesthetic delivery system by Magill,
popularly known as the “Magill’s circuit”. Introduction of cyclopropane in
1929 and cuffed endotracheal tube in 1931; prompted Waters to develop the “to
and fro” canister and use it for closed system anaesthesia with cyclopropane.
In 1936, Brian Sword introduced the circle system. The Ayre’s T-piece was
introduced in 1937, EMO inhaler in 1941 and Minnitt’s ‘gas and air’
apparatus with demand valve in 1949.
With
the introduction of many breathing systems attempts were made to classify them
in the 50’s and 60’s, but lack of a proper definition, lead to more
confusion than clarity.
DEFINITION:
A
breathing system is defined as an assembly of components which connects the
patient’s airway to the anaesthetic machine creating an artificial atmosphere,
from and into which the patient breathes.
It
primarily consists of
a)
A fresh gas entry port/delivery tube through which the gases are delivered from
the machine
to the systems;
b)
A port to connect it to the patient’s airway;
c)
A reservoir for gas, in the form of a bag or a corrugated tube to meet the peak
inspiratory flow requirements;
d)
An expiratory port/valve through which the expired gas is vented to the
atmosphere;
e)
A carbon dioxide absorber if total rebreathing is to be allowed and
f)
Corrugated tubes for connecting these components.
Flow
directing valves may or may not be used.
REQUIREMENTS
OF A BREATHING SYSTEM:
The
components when assembled should satisfy certain requirements, some essential
and others desirable.
Essential:
The
breathing system must
a)
deliver the gases from the machine to the alveoli in the same concentration as
set and in the shortest possible time;
b)
effectively eliminate carbon-dioxide;
c)
have minimal apparatus dead space; and
d)
have low resistance.
Desirable:
The
desirable requirements are
a)
economy of fresh gas;
b)
conservation of heat;
c)
adequate humidification of inspired gas;
d)
Light weight;
e)
convenience during use;
f)
efficiency during spontaneous as well as controlled ventilation (Efficiency is
determined in terms of CO2 elimination and fresh gas utilization);
g)
adaptability for adults, children and mechanical ventilators;
h) provision to reduce theatre pollution.
CLASSIFICATION
OF BREATHING SYSTEMS:
One
will realize the reason for the failure of the attempts at classification in the
50’s to 60’s, if this definition and requirements are taken into account.
There are numerous classifications of breathing systems according to the whims
and fancy of the person classifying. Many of them are irrelevant as they do not
define a breathing system. Different authors classified the same system under
different headings, adding to confusion1.
McMohan in 1951 classified them as open, semiclosed and closed taking the
level of rebreathing into account. It as follows:
|
Open |
no rebreathing |
| Semiclosed | partial rebreathing |
| Closed | total rebreathing |
Dripps
et al have classified them as Insufflation, Open, Semiopen, Semiclosed and
Closed taking into account the presence or absence of Reservoir, Rebreathing, CO2
absorption and Directional valves1. The ambiguity of the terminology
used as open, semi open, semi closed and closed allowed inclusion of apparatus
that are not breathing systems at all into the classification.
To
overcome this problem Conway2 suggested that a functional
classification be used and classified according to the method used for CO2
elimination as:
1.
Breathing systems with CO2 absorber and
2.
Breathing systems without CO2 absorber.
| BREATHING SYSTEMS WITHOUT CO2 ABSORPTION. | I BREATHING SYSTEMS WITH CO2 ABSORPTION. |
|
Unidirectional
flow: a) Non rebreathing systems. B) Circle systems.
|
Unidirectional
flow Circle system with absorber.
|
|
Bi-directional
flow: a) Afferent reservoir systems. Mapleson A Mapleson B Mapleson C Lack’s system. B) Enclosed afferent reservoir systems Miller’s (1988) c) Efferent reservoir systems Mapleson D Mapleson E Mapleson F Bain’s system d) Combined systems Humphrey ADE
|
Bi-directional
flow To and Fro system.
|
BREATHING
SYSTEMS WITHOUT CO2 ABSORPTION
|
|
a)
Nonrebreathing systems. They use non rebreathing valves and there is no mixing of
fresh gas and the expired gas.
Functional
analysis:
When the patient takes a breath, or if the reservoir bag is squeezed, the
inspiratory unidirectional valve opens and the gases flow into the patient’s
lungs(Fig.1). The expiratory
unidirectional valve closes the expiratory port during spontaneous breathing. The inspiratory unidirectional valve itself closes the
expiratory port during controlled ventilation.
At the start of expiration, the inspiratory unidirectional valve returns
back to position and expiration takes place through the expiratory port, opening
the expiratory valve.
The
fresh gas flow (FGF) should be equal to the minute ventilation (MV)
of the patient. These systems satisfy all four essential requirements, but are
not very popular because of the following reasons:
1)
Fresh gas flow has to be constantly adjusted and is not economical.
2)
There is no humidification of inspired gas.
3)
There is no conservation of heat.
4)
They are not convenient as the bulk of the valve has to be positioned near the
patient.
5)
The valves can malfunction due to condensation of moisture and lead to
complications.
B)
Circle Systems: These
systems are designed with a CO2 absorber as an essential component of
the system. To use it without absorber is uneconomical as it needs a FGF more
than the alveolar ventilation.
The
effect of arrangements of various components in the effective elimination of CO2
and fresh gas economy when used with high flows were analysed by Egar and Ethans7.
Detailed discussion on this is beyond the purview of this review.
However, some aspects of this is discussed under the section, Circle
system with absorber.
Bi-Directional
Flow:
Systems
with bi-directional flow are extensively used. These systems depend on the FGF
for effective elimination of CO2. Understanding these systems is most
important as their functioning can be manipulated by changing
parameters like Fresh gas flow, alveolar ventilation, apparatus dead
space, etc. We will analyze these
in detail.
Fresh
Gas Supply; Fresh gas flow (FGF) forms one of the essential requirements
of a breathing system. If there is
no FGF into the system, the patient will get suffocated. If the FGF is low, most systems do not eliminate
carbon-dioxide effectively, and if there is an excess flow there is wastage of
gas. So, it becomes imperative to
specify optimum FGF for a breathing system for efficient functioning.
If
the system has to deliver a set concentration in the shortest possible time to
the alveoli, the FGF should be
delivered as near the patient’s airway as possible.
Elimination
Of Carbon-Dioxide: The following may be taken as an example for better
understanding of CO2 elimination by the bi-directional flow systems.
Normal production of carbon-dioxide in a 70 kg adult is 200 ml per minute
and it is eliminated through the lungs. Normal
end-tidal concentration of carbon-dioxide is 5%.
Hence, for eliminating 200 ml of carbon-dioxide as a 5% gas mixture, the
alveolar ventilation has to be:
200
x 100 = 4,000 ml.
5
This 4000 ml or 4 litres is the normal alveolar ventilation. Any breathing system connected to an adult’s airway should provide a minimum of 4 litres per minute of carbon-dioxide free gas to the alveoli for eliminating carbon-dioxide. If the alveolar ventilation becomes less than 4 litres per minute, it would lead to hypercarbia. If the alveoli are ventilated with 5 litres/minute of a gas containing 1% carbon-dioxide, or 8 litres/minute of a gas containing 2.5% carbon-dioxide, it could still eliminate 200 ml of carbon-dioxide per minute from the alveoli. It may be construed as 4 litres of CO2 free gas and 1 litre of gas with 5% CO2 in the first instant and as 4 litres of CO2 free gas and 4 litres of gas with 5% CO2 in the second instant. In effect, 4 litres of alveolar ventilation with CO2 free gas is provided in both cases.
Apparatus
Dead Space: It is the volume of the breathing system from the patient-end
to the point up to which, to and fro movement of expired gas takes place.
|
|
In
an afferent reservoir system with adequate FGF, the apparatus dead space extends
up to the expiratory valve positioned near the patient (fig.2).
If the FG enters the system near the patient-end as in an efferent reservoir system, the dead space extends upto the point of FG entry. In systems where inspiratory and expiratory limbs are separate, it extends upto the point of bifurcation. The dynamic dead space will depend on the FGF and the alveolar ventilation. The dead space is minimal with optimal FGF. If the FGF is reduced below the optimal level, the dead space increases and the whole system will act as dead space if there is no FGF. Increasing the FGF above the optimum level will only lead to wastage of FG.
Sub-Classification
Of Bi-Directional Flow Systems:
Mapleson8 did a theoretical analysis of the fresh gas requirements of the semiclosed systems available at that time. It is only proper to refer to it as Mapleson systems as he gave a nomenclature as A, B, C, D and E for easy identification as per their construction. For better understanding of the functional analyses, they have been classified as:
| 1 | Afferent reservoir system (ARS). |
| 2 | Enclosed afferent reservoir systems (EARS). |
| 3 | Efferent reservoir systems (ERS). |
| 4 | Combined systems. |
The
afferent limb is that part of the breathing system which delivers the fresh gas
from the machine to the patient. If
the reservoir is placed in this limb as in Mapleson A, B, C and Lack’s
systems, they are called afferent reservoir systems (ARS).
The
efferent limb is that part of the breathing system which carries expired gas
from the patient and vents it to the atmosphere through the expiratory
valve/port. If the reservoir is
placed in this limb as in Mapleson D, E, F and Bain systems, they are called
efferent reservoir systems (ERS).
Enclosed afferent reservoir system has been described by Miller and Miller.
AFFERENT
RESERVOIR (AR) SYSTEMS
The
Mapleson A, B and C systems have the reservoir in the afferent limb, and do not
have an efferent limb (Fig.3). Lack system has an afferent limb reservoir and an
efferent limb through which the expired gas traverses before being vented into
the atmosphere (Fig.4). This limb is coaxially placed inside the afferent limb.
|
|
Mapleson8 has analysed these bi-directional flow systems using mathematical calculations. He made a few basic assumptions while analyzing breathing systems. These are
(1) Gases move enbloc. They maintain their identity as fresh gas, dead space gas and alveolar gas. There is no mixing of these gases.
(2) The reservoir bag continues to fill up, without offering any resistance till it is full.
(3) The expiratory valve opens as soon as the reservoir bag is full and the pressure inside the system goes above atmospheric pressure.
(4)
The valve remains open throughout the expiratory phase without offering any
resistance to gas flow and closes at the start of the next inspiration.
Mapleson
‘A’/Magill’s system:
Functional
analysis:
Spontaneous breathing:
The system is filled with fresh gas before connecting to the patient.
When the patient inspires, the fresh gas
from the machine and the reservoir bag flows to the patient, and as a
result the reservoir bag collapses (Fig.5a). During expiration, the FG
continues to flow into the system and fill the reservoir bag. The expired gas, initial part of which is the dead space gas,
pushes the FG from the corrugated tube into the reservoir bag and collects
inside the corrugated tube (Fig.5b).
|
|
As soon as the reservoir bag is full, the expiratory valve opens and the alveolar gas is vented into the atmosphere (Fig.5c). During the expiratory pause, alveolar gas that had come into the corrugated tube is also pushed out through the valve, depending on the FGF. The system is filled with only fresh gas and dead space gas at the start of the next inspiration when FGF is equal to the alveolar ventilation (Fig.5d). The entire alveolar gas and dead space gas is vented through the valve and some FG also escapes, if the FGF is higher than the minute ventilation. Some amount of alveolar gas will remain in the system and lead to rebreathing with a FGF less than the alveolar ventilation. This has been confirmed theoretically and experimentally by many investigators8,9. The system functions at maximum efficiency, when the FGF equals the alveolar ventilation and the dead space gas (which has not taken part in gas exchange) is allowed to be rebreathed and utilized for alveolar ventilation.
Controlled ventilation: To facilitate IPPV the expiratory valve has to be partly closed. During inspiration, the patient gets ventilated with FG and part of the FG is vented through the valve (Fig.6a) after sufficient pressure has developed to open the valve. During expiration, the FG from the machine flows into the reservoir bag and all the expired gas (i.e., dead space gas and alveolar gas) flows back into the corrugated tube till the system is full (Fig.6b). During the next inspiration the alveolar gas is pushed back into the alveoli followed by the FG. When sufficient pressure is developed, part of the expired gas and part of the FG escape through the valve (Fig.6c). This leads to considerable rebreathing, as well as excessive waste of fresh gas. Hence these systems are inefficient for controlled ventilation.
|
|
Lack’s
system:
This
system functions like a Mapleson A system both during spontaneous and controlled
ventilation. The only difference is
that the expired gas instead of getting vented through the valve near the
patient, is carried by an efferent tube placed coaxially and vented through the
valve placed near the machine end (Fig.4).
This facilitates easy scavenging of expired gas.
|
|
Mapleson
B & C systems:
In
order to reduce the rebreathing of alveolar gas and to improve the utilization
of FG during controlled ventilation, the FG
entry was shifted near the patient(Fig.3).
This allows a complete mixing of FG and expired gas.
The end result is that these systems are neither efficient during
spontaneous nor during controlled ventilation.
ENCLOSED
AFFERENT RESERVOIR (EAR) SYSTEMS
This
has been described by Miller & Miller10.
The system consisted of a Mapleson A system enclosed within a non
distensible structure (Fig.7a). It
may also be constructed by enclosing the reservoir bag alone in a bottle and
connecting the expiratory port to the bottle with a corrugated tube and a one
way valve (Fig.7b). To the bottle
is also attached a reservoir bag and a variable
orifice for providing positive pressure ventilation.
|
|
Functional
analysis: During
spontaneous ventilation, the gas is vented from the system in a manner which is
identical to the Mapleson A system. In
this mode the variable orifice is kept
widely open to allow free communication to the atmosphere. In controlled
ventilation the reservoir bag ‘B’
is squeezed intermittently and the variable
orifice is partly closed to allow building up of pressure in the bottle.
The pressure thus developed (1) closes the expiratory valve, (2) squeezes
the enclosed afferent reservoir and the patient gets ventilated. The expiration
takes place in a manner similar to that described during spontaneous ventilation
when the pressure is released in reservoir ‘B’,.
Hence this system should function efficiently during spontaneous and
controlled ventilation with a FGF equivalent to alveolar ventilation. The fresh
gas requirement and the utilization of this system has been investigated by a
group of investigators from Manchester11-13 and a group from Wales14
under the guidance of Mapleson. They have reported varying figures for
utilization as 82%, 93% and 74% respectively11,12,14.
The reasons for this lesser percentage of utilization have been quoted as
faulty methodology for calculation15, resistance offered by the
reservoir bag and tubing and early opening of the unidirectional valve during
expiration14 etc. Though the fresh gas requirement is higher than the
alveolar ventilation in this system as shown by the above studies, it is still
more efficient than the Bain system for controlled ventilation.
EFFERENT
RESERVOIR (ER) SYSTEMS:
|
|
|
|
Bain system |
|
|
The Mapleson D, E, F and Bain systems have a 6 mm tube as the afferent limb that supplies the FG from the machine. The efferent limb is a wide-bore corrugated tube to which the reservoir bag is attached and the expiratory valve is positioned near the bag. In Mapleson E system, the corrugated tube itself acts as the reservoir (Fig.8). In Bain system, the afferent and efferent limbs are coaxially placed (Fig.9).
All
these ER systems are modifications of Ayre’s T-piece.
This consists of a light metal tube 1 cm in diameter, 5 cm in length with
a side arm (Fig.10). Used as such,
it functions as a non-rebreathing system. Fresh
gas enters the system through the side arm and the expired gas is vented into
the atmosphere and there is no rebreathing.
The dead space is minimal as it is only up to the point of FG entry and
elimination of CO2 is achieved by breathing into the atmosphere.
FGF equal to peak inspiratory flow rate of the patient has to be used to
prevent air dilution.
|
|
In
an attempt to reduce FGF requirements, ER systems are constructed with
reservoirs in the efferent limb. The
functioning of all these systems are similar.
These systems work efficiently and economically for controlled
ventilation as long as the FG entry and the expiratory valve are separated by a
volume equivalent to atleast one tidal volume of the patient.
They are not economical during spontaneous breathing.
Spontaneous respiration:
The breathing system should be filled with FG before connecting to the
patient. When the patient takes an inspiration, the FG from the machine, the
reservoir bag and the corrugated tube flow to the patient (Fig.11a).
During expiration, there is a continuous FGF into the system at the
patient end. The expired gas gets
continuously mixed with the FG as it flows back into the corrugated tube and the
reservoir bag (Fig.11b). Once the
system is full the excess gas is vented to the atmosphere through the valve
situated at the end of the corrugated tube near the reservoir bag.
During the expiratory pause the FG
continues to flow and fill the proximal portion of the corrugated tube while the
mixed gas is vented through the valve (Fig.11c).
During the next inspiration, the patient breaths FG as well as the mixed
gas from the corrugated tube (Fig.11d). Many
factors influence the composition of the inspired mixture.
They are FGF, respiratory rate, expiratory pause, tidal volume and CO2
production in the body. Factors
other than FGF cannot be manipulated in a spontaneously breathing patient.
It has been mathematically calculated and clinically proved8,16
that the FGF should be atleast 1.5 to 2 times the patient’s minute ventilation
in order to minimise rebreathing to acceptable levels.
|
|
Controlled
ventilation:
To facilitate intermittent positive pressure
ventilation, the expiratory valve has to be partly closed so that it opens only
after sufficient pressure has developed in the system.
When the system is filled with fresh gas, the patient gets ventilated
with the FGF from the machine, the corrugated tube and the reservoir bag
(Fig.12a). During expiration, the
expired gas continuously gets mixed with the fresh gas that is flowing into the
system at the patient end. During
the expiratory pause the FG continues to enter the system and pushes the mixed
gas towards the reservoir (12B). When the next inspiration is initiated, the patient gets
ventilated with the gas in the corrugated tube i.e., a mixture of FG, alveolar
gas and dead space gas (Fig.12c). As
the pressure in the system increases, the expiratory valve opens and the
contents of the reservoir bag are discharged into the atmosphere.
|
|
Factors
that influence the composition of gas mixture in the corrugated tube with which
the patient gets ventilated are the same as for spontaneous respiration namely
FGF, respiratory rate, tidal volume and pattern of ventilation.
The only difference is that these parameters can be totally controlled by
the anaesthesiologist and do not depend on the patient. Using a low respiratory
rate with a long expiratory pause and a high tidal volume, most of the FG could
be utilized for alveolar ventilation without wastage.
Analyzing
the performance of these systems during controlled ventilation, two
relationships have become evident. 1)
When FGF is very high the PaCO2 becomes ventilation dependent (as
during spontaneous respiration). 2)
When the minute volume exceeds the FGF substantially, the PaCO2 is
dependent on the FGF17. Combining
these influences a graph can be constructed as shown in Fig.13.
An infinite number of combinations of FGF and minute ventilation can be
chosen to achieve a desired PaCO2. One can use a high FGF and a
normal minute volume of 70 ml/kg to achieve a normal PaCO2 of 40 mm
Hg. This is uneconomical and leads
to low humidity and heat loss. Alternately,
a FGF equivalent to the predicted minute volume i.e., 70 ml/kg can be chosen and
the patient ventilated with at least twice the predicted minute volume i.e. 140
ml/kg. Here a deliberate controlled
rebreathing is allowed in order to maintain normal PaCO2 along with
high humidity, less heat loss and greater economy of fresh gas.
Combinations between these two extremes can also be used.
It is important to remember that using a low FGF with normal minute
ventilation, can lead to hypercarbia; a moderate FGF and hyperventilation, can
lead to hypocarbia.
COMBINED
SYSTEMS
To
over come the difficulties of changing the breathing systems for different modes
of ventilation, Humphrey designed a system called Humphrey ADE18,
with two reservoirs, one in the afferent limb and the other in the efferent
limb. While in use, only one reservoir will be in operation and the
system can be changed from ARS to ERS by changing the position of a lever.
It can be used for adults as well as children.
The functional analysis is the same as Mapleson A in ARS mode and as Bain
in ERS mode. It is not yet widely
used.
BREATHING
SYSTEMS WITH CO2 ABSORPTION
Systems
so far described have relied on FGF for effective elimination of CO2.
Any desire to economize on FGF by allowing a total rebreathing, should be
accompanied by removal of the expired CO2 by chemical
absorption using sodalime or baralyme.
The systems designed for these purpose are again classified as:
Unidirectional
flow.
-Circle
system.
Bi-directional
flow.
-To and fro system.
The essential components of the circle system are, (1) a sodalime canister, (2) Two unidirectional valves, (3) Fresh gas entry, (4) Y-piece to connect to the patient, (5) Reservoir bag (6) a relief valve and (7) low resistance interconnecting tubing. The arrangement of the components is shown in fig.14. For efficient
|
|
functioning
of the system the following criteria should fulfilled.
(1) There should be two unidirectional valves on either side of the
reservoir bag, (2) Relief valve should be positioned in the expiratory limb
only, (3) The FGF should enter the system proximal to the inspiratory
unidirectional valve.
Functional
analysis: During inspiration the FG along with the CO2 free
gas in the reservoir bag flow through the inspiratory limb and inspiratory
unidirectional valve to the patient. No
flow takes place in the expiratory limb as the expiratory unidirectional valve
is closed by back pressure transmitted to the valve. During expiration the inspiratory unidirectional valve closes
and the expired gas flows through the expiratory unidirectional valve in the
expiratory limb to the sodalime canister and to the reservoir bag.
The CO2 is absorbed in the canister.
The FGF from the machine continues to fill the reservoir bag.
When the reservoir is full the relief valve opens and the excess gas is
vented to atmosphere. By selecting
a suitable position for the relief valve, the expired gas can be selectively
vented when the FGF is more than the alveolar ventilation. To facilitate
controlled ventilation the relief valve has to be partly closed and the excess
gas is vented during inspiration. The gas flow pattern is similar to that
described above.
The
advantages and disadvantages of the various arrangements of the components were
analyzed by Eger and Ethans7. The
relative positions of the components of the circle system are of particular
importance to the functioning of the system only when the FGF is high, the gas
components of the system unmixed and CO2 absorber not used.
When the FGF is reduced below the alveolar ventilation, the CO2
absorber is a must as the gas in the system become more uniformly mixed, and the
relative position of the system’s components become less important
Totally
closed system:
The
systems with CO2 absorption can be used in a completely closed mode.
After a period of approximately 10-20 minutes breathing with high inflow
of fresh gas for denitrogenation, the expiratory valve is closed.
The FGF is then adjusted to meet only the patients basal oxygen
requirements together with anaesthetic. A
number of advantages have been demonstrated for totally closed systems.
A)
Economy: The FGF could be reduced
to as low as 250 - 500 ml of oxygen. The
consumption of Halothane/Isoflurane has been found to be around 3.5 ml/hour19.
b)
Humidification: In the completely
closed system, once the equilibrium has been
established, the inspired gas will be fully saturated with water vapour20.
C)
Reduction of heat loss: In addition
to conserving water the totally closed system will also conserve heat. The CO2
absorption is an exothermic reaction and the system may actively assist in
maintaining body temperature.
D)
Reduction in atmospheric pollution: Once
the expiratory valve has been closed, no anaesthetic escapes, except for the
small percutaneous loss from the patient.
E)
Control of anaesthesia: It is
possible to compute the time course of uptake of anaesthetic in a patient of
known size and add the appropriate quantity of the anaesthetic to the circuit at
a rate decreasing in a manner calculated to maintain a constant alveolar
concentration21. In
practice an alveolar concentration of about 1.3 x MAC is found to be suitable.
The
technique has several potential disadvantages.
i)
A greater knowledge of uptake and distribution is required to master closed
circuit anaesthesia.
ii)
Inability to alter any concentration quickly.
iii)
Real danger of hypercapnia may result from, a) an inactive absorber, B)
incompetent unidirectional valves and c) incorrect use of absorber bypass.
Bi-directional
flow systems
The
Waters to and fro system is valveless and conveniently portable. It has been
widely used in the past and now is only of historical importance.
The reader may refer to any standard text book for furtherdetails.
References
1.
Dorsch, J.A., Dorsch,S.E. Breathing
systems II. In understanding
anaesthesia equipment. 2nd ed.
1984. Williams & Wilkins.
2.
Conway, C.M.: Anaesthetic breathing
systems. British Journal of Anaesthesia. 1985, 57, 649-57.
3.
Miller, D.M.: Breathing systems for use in anaesthesia.
Evaluation using a physical lung model and classification.
British Journal of Anaesthesia. 1988, 60, 555-64.
4.
Miller DM. An enclosed efferent afferent reservoir system: The Maxima.
Anaesthesia and intensive care 1995; 23:284-91.
5.
Miller DM. Breathing systems reclassified. Anaesthesia and intensive care 1995;
23: 281-83.
6.
Miller DM, Palm A. Comparison in spontaneous ventilation of the Maxima with the
Humphrey ADE breathing system and between four methods for detecting
rebreathing. Anaesthesia and intensive care 1995; 23: 296-01.
7.
Eager EI., Ethans, C.T.: The
effects of inflow, overflow and valve placement
on economy of circle system. Anesthesiology,
1968, 29, 93-100.
8.
Mapleson WW. The elimination of rebreathing in various semiclosed anaesthetic
systems. British journal of Anaesthesia; 1954;26: 323-32.
9.
White, DC, Calkins, J. Anaesthetic
machine and breasting apparatus. In:
Nunn, Utting and Brown eds. General anaesthesia.
5th ed.. Butterworths, London 1989:428-56
10. Miller, D.M., Miller, J.C.:
Enclosed afferent reservoir breathing systems. Description and clinical evaluation. British Journal of Anaesthesia. 1988, 60, 469-75.
11.
Droppert PM, Meakin G, Beatty PCW, Mortimer AJ, Healy TEJ. Efficiency of a new
afferent reservoir breathing system during controlled ventilation. British
Journal of Anaesthesia 1991; 66: 638-42.
12.
Meakin G, Jennings AD, Beatty PCW, Healy TEJ. Fresh gas requirements of an
enclosed afferent reservoir breathing system
during controlled ventilation in children. British Journal of Anaesthesia
1992; 68: 43-47.
13.
Beatty PCW, Meakin G, Healy TEJ. Fractional delivery of fresh gas: a new index
of the efficiency of semiclosed breathing systems. British Journal of
Anaesthesia 1992; 68: 474-77.
14.
Tham EJ, Davies R, Slade JM, Mapleson WW. Efficiency of breathing systems A and
D in the Carden Ventmasta ventilator. British Journal of Anaesthesia 1993; 71:
741-46.
15.
Ravishankar M, Chatterjee S. Fractional utilisation of fresh gas by breathing
systems without carbon dioxide absorption. British Journal of Anaesthesia 1993;
71: 706-07
16.
Ward CS. In: Anaesthetic equipment. Physical
principles and maintenance; W.B.Saunders, London; 2nd ed. 1985.
17.
Rose DK, Froese AB. The regulation of PaCO2 during controlled
ventilation of children with a T-piece. Canadian Anaesthetists Society Journal
1979; 26: 104-13.
18.
Humphrey D. A new anaesthetic breathing system combining Mapleson A,D and E
principles. Anaesthesia 1983; 38: 361-72
19.
Baum JA, Aitkenhead AR. Low flow anaesthesia. Anaesthesia 1995;
50(supplement):37-44.
20.
Kleemann, P.P. Humidity of
anaesthetic gases with respect to low flow anaesthesia.
Anaesthesia and Intensive Care, 1994, 22 (4), 396-408.
21.
Lowe HJ, Ernst EA. The quantitative practice of anaesthesia, Use of closed
circuit. Baltimore. Williams & wilkins, 1981.